HydroTech ZS Consulting Rev 7 (2023)
Water Quality Monitoring Equipment 
 
This app was created by HydroTech to provide our customers with a quick access to calculations and conversions that they use on a daily basis while calibrating and maintaining their water quality monitoring equipment:
Temperature: Always check the temperature sensor against an NIST thermometer
  
Oxygen Solubility:  In order to calibrate dissolved oxygen sensors to DO (mg/L), the end user needs to determine the solubility of oxygen in the water first. Input the current uncompensated barometric pressure (mmHg). Input current temperature reading from the multiprobe thermometer (°C). Input the current SpConductivity value of the water used for calibration (µs/cm). Click "Enter", "Tab", "Go" or just click on the black field to display the results
 
pH Buffer Temp.Correction: All pH buffers values are rated to 25 °C. When using a buffer that has a temperature other than 25 °C, the end user needs to apply a correction to the buffer value based on its current temperature. Select pH buffer 4,7 or 10 column. Input the current buffer temperature (°C). Click "Enter", "Tab", "Go" or just click on the black field to display the results. 
 
pH Sensor Health: Input current pH (mV) readings for each pH buffer. Click "Enter", "Tab", "Go" or just click on the black field to display the results and the health status for your pH sensor. 
 
ORP Zobell Temp. Correction: The ORP Zobell solution value is rated at 25 °C. When using a Zobell solution that has a temperature other than 25 °C, the end user needs to apply a correction to the solution value based on its current temperature. Inpute the Zobell solution value at 25 °C. Input the current Zobell solution temperature (°C). Click "Enter", "Tab", "Go" or just click on the black field to display the results. 
 
SpCond to Salinity Conversion: Input the SpConductivity value (µs/cm). Click "Enter", "Tab", "Go" or just click on the black field to display the results. The Salinity conversion is based on USGS Water-Supply Paper 2311
 
Salinity to Chloride Conversion: Input the salinity value (ppt). Click "Enter", "Tab", "Go" or just click on the black field to display the results.
 
S4000 NTU Formazin Dilution: The Formazin is the recommended standard for calibrating the turbidity sensor. The 4000 (NTU) standard will need to be diluted to lower values by the end user. Input the value of the standard needed (NTU). Input the total standard volume needed (ml). Click "Enter", "Tab", "Go" or just click on the black field to display the results.
 
NO3 Dilution: Start with a 50 mg/l NO3 standard. Input NO3 standard value needed (mg/l), Input Total standard volume needed (ml). Click "Enter", "Tab", "Go" or just click on the black field to display the results.
 
TDG Percent: Input TDG% value (mmHg). Input Barometric Pressure (mmHg). Click "Enter", "Tab", "Go" or just click on the black field to display the results.
 
Actual B.P. (mmHg):  When you obtain the Barometric Pressure (B.P.) measurement from a weather station they usually report the corrected B.P to sea level. You need to uncorrect that value based on your current altitude (Actual B.P.). Input the corrected B.P. (mmHg) or (inHg). Input your current altitude (f). Click "Enter", "Tab", "Go" or just click on the black field to display the results.
 
NO3-N to NO3: Input the N03 value in mg/l-N. Click "Enter", "Tab", "Go" or just click on the black field to display the converted value to NO3-N and vice versa.  
http://www.hydrotechzs.com
Telephone: 512-846-2893
 Rev7- 2023
Zak Sihalla
            Status    
    Sonde accuracy °C:   0.2        
                 
    NIST Therm. accuracy °C:          
                 
    Current NIST Therm. reading °C:          
                 
    Current sonde reading °C:          
                 
    Calculated accuracy °C:        
                 
 
Date:        
               
Name:        
               
SN:        
               
Notes:        
               
Recommended setup for checking the temperature sensor at room temperature:
1- Use tap water that was stored overnight in the same room you are doing the temperature check “stable water.”
2- Set the sonde vertically with the sensors facing up and remove the storage cup cap.
3- Fill the storage cup until all sensors are submerged in the “stable water.”
4- Insert the NIST temperature sensor into the storage cup. Make sure the NIST sensor tip is touching the sonde temperature sensor tip.
5- Wait for the readings to stabilize before taking a measurement.
 
http://www.hydrotechzs.com    
Copyright © 2023 HydroTech ZS Consulting. All rights reserved     
                     
Current B.P. (mmHg):      
Current temperature (°C):      
Current SpConductivity (µs/cm):      
 
Oxygen solubility (mg/l):      
                     
                     
  Date/Time:    
  Name:    
  SN:    
  Notes:    
         
         
                     
In order to calibrate dissolved oxygen sensors to DO (mg/L), the end user needs to determine the solubility of oxygen in the water first. Input the current uncompensated barometric pressure (mmHg). Input current temperature reading from the multiprobe thermometer (°C). Input the current SpConductivity value of the water used for calibration (µs/cm). Click "Enter", "Tab", "Go" or just click on the black field to display the results  
http://www.hydrotechzs.com  
Copyright © 2023 HydroTech ZS Consulting. All rights reserved  
                   
Buffer value @ 25 (°C):   4   7   10      
Current buffer temp.(°C):            
 
 
 
Buffer value @ current temp.:            
                   
                   
                   
Buffer value @ 25 (°C):   6   6.86          
Current buffer temp.(°C):              
 
 
 
Buffer value @ current temp.:              
                   
                   
Date/Time:      
Name:      
SN:      
Notes:      
       
       
                 
All pH buffers values are rated to 25 °C. When using a buffer that has a temperature other than 25 °C, the end user needs to apply a correction to the buffer value based on its current temperature. Select pH buffer 4,7 or 10 column. Input the current buffer temperature (°C). Click "Enter", "Tab", "Go" or just click on the black field to display the results.     
http://www.hydrotechzs.com    
Copyright © 2023 HydroTech ZS Consulting. All rights reserved    
               
Buffers:   7   10   4  
               
 pH (mV)::        
 
 
Slope %:          
               
Status:         
               
               
               
Date/Time:
Name:
SN:
Notes:
   
   
 
 Input current pH (mV) readings for each pH buffer. Click "Enter", "Tab", "Go" or just click on the black field to display the results and the health status for your pH sensor.  
http://www.hydrotechzs.com
Copyright © 2023 HydroTech ZS Consulting. All rights reserved 
               
                             
Zobell @ 25 °C (mV):      
Current Zobell temperature (°C):      
 
Zobell value @ current temperature (mV):      
                 
 Standard Methods for the Examination of Water and Wastewater  
Eh, V = 0.229 - 0.0022 (T -25)      
 
Date/Time:  
Name:  
SN:  
Notes:  
     
       
         
The ORP Zobell solution value is rated at 25 °C. When using a Zobell solution that has a temperature other than 25 °C, the end user needs to apply a correction to the solution value based on its current temperature. Inpute the Zobell solution value at 25 °C. Input the current Zobell solution temperature (°C). Click "Enter", "Tab", "Go" or just click on the black field to display the results.   
http://www.hydrotechzs.com  
Copyright © 2023 HydroTech ZS Consulting. All rights reserved   
               
 Enter value then click "Tab" or Touch the black field to display result
SpConductivity (µs/cm):  
Salinity (ppt):  
Salinity conversion based on USGS Water-Supply Paper 2311
               
 
Date/Time:
Name:
SN:
Notes:
   
   
 
 Input the SpConductivity value (µs/cm). Click "Enter", "Tab", "Go" or just click on the black field to display the results. The Salinity conversion is based on USGS Water-Supply Paper 2311
http://www.hydrotechzs.com
Copyright © 2023 HydroTech ZS Consulting. All rights reserved 
                   
Enter value then click "Tab" or Touch the black field to display result  
Salinity (ppt):      
Chloride (ppt):      
                   
                   
Date/Time:    
Name:    
SN:    
Notes:    
       
       
                   
   Input the salinity value (ppt). Click "Enter", "Tab", "Go" or just click on the black field to display the results.  
http://www.hydrotechzs.com  
Copyright © 2023 HydroTech ZS Consulting. All rights reserved   
                   
                   
Standard value needed (NTU):     
Total standard volume needed (ml):    
Volume of 4000 NTU concentrate to use (ml):     
Volume of DI water to use (ml):     
                   
                   
Date/Time:    
Name:    
SN:    
Notes:    
       
       
                   
The Formazin is the recommended standard for calibrating the turbidity sensor. The 4000 (NTU) standard will need to be diluted to lower values by the end user. Input the value of the standard needed (NTU). Input the total standard volume needed (ml). Click "Enter", "Tab", "Go" or just click on the black field to display the results.
http://www.hydrotechzs.com
Copyright © 2023 HydroTech ZS Consulting. All rights reserved 
            NO3   Nox  
Standard value needed (mg/l):        
Total standard volume needed (ml):        
Volume of 50 mg/l  concentrate to use (ml):        
Volume of DI water to use (ml):        
                       
                       
Date/Time:      
Name:      
SN:      
Notes:      
         
         
                       
Start with a 50 mg/l NO3 standard. Input NO3 standard value needed (mg/l), Input Total standard volume needed (ml). Click "Enter", "Tab", "Go" or just click on the black field to display the results.
http://www.hydrotechzs.com  
Copyright © 2023 HydroTech ZS Consulting. All rights reserved   
                   
TDG (MMhG):      
Barometric Pressure (mmHg):      
TDG %:      
                 
 
Date/Time:  
Name:  
SN:  
Notes:  
     
     
 
Input TDG% value (mmHg). Input Barometric Pressure (mmHg). Click "Enter", "Tab", "Go" or just click on the black field to display the results  
 
http://www.hydrotechzs.com    
Copyright © 2023 HydroTech ZS Consulting. All rights reserved     
                           
Corrected B.P (mmHg or inHg):   mmHg inHg  
Altitude (ft):                
Actual B.P. (mmHg):                
                           
                           
Date/Time:            
Name:            
SN:            
Notes:            
               
               
                           
When you obtain the Barometric Pressure (B.P.) measurement from a weather station they usually report the corrected B.P to sea level. You need to uncorrect that value based on your current altitude (Actual B.P.). Input the corrected B.P. (mmHg) or (inHg). Input your current altitude (f). Click "Enter", "Tab", "Go" or just click on the black field to display the results.
                           
http://www.hydrotechzs.com
Copyright © 2023 HydroTech ZS Consulting. All rights reserved 
                     
NO3-N (mg/l-N):      
NO3 (mg/l):      
                     
 
Date/Time:  
               
Name:  
                 
SN:  
                 
Notes:  
     
     
 
 Input the N03 value in mg/l-N. Click "Enter", "Tab", "Go" or just click on the black field to display the converted value to NO3-N and vice versa.  
 
http://www.hydrotechzs.com
Copyright © 2023 HydroTech ZS Consulting. All rights reserved