NPC Service Calculator

Reference voltage & absorbance validation for NICO & OPUS nitrate probes

NICO IN ZERO SOLUTION

Submerge the lenses in ultra pure water then input values below.

ParameterInput valueStatusComments
SQISensor Quality Index
RefD / Reference Diode
RefA / CH1 212 nm
RefB / CH2 254 nm
RefC / CH3 360 nm
NICO IN MEDIUM

Submerge the lenses in the medium to be measured then input values below.

ParameterInput valueStatusComments
RefD / Reference Diode
RefA / CH1 212 nm
RefB / CH2 254 nm
RefC / CH3 360 nm
RefA < RefB < RefC
🔬Enter values above to validate the NICO probe.
OPUS IN ZERO SOLUTION

Submerge the lenses in ultra pure water then input values below.

ParameterInput valueStatusComments
SQISensor Quality Index
ABS 210 AUNutrients
ABS 254 AUOrganic
ABS 360 AUTurbidity
ABS 210 = 254 = 360
OPUS IN MEDIUM

Submerge the lenses in the medium to be measured then input values below.

ParameterInput valueStatusComments
SQISensor Quality Index
ABS 210 AUNutrients
ABS 254 AUOrganic
ABS 360 AUTurbidity
ABS 210 > 254 > 360
🔬Enter values above to validate the OPUS probe.